1. Introduction
This standard operating procedure (SOP) describes a procedure for acid digestion of samples using the Multiwave GO Plus microwave system. A known weight of the sample is placed in the digestion vessel and digestion acid added. The digestion vessel is capped and subjected to microwave heating within a sealed pressurized chamber to achieve a high digestion temperature, then cooled down to room temperature. The digest liquor is quantitatively transferred into a plastic test tube and diluted with water to a known volume.
2. Scope
This procedure may be applied to any organic sample that is amenable to acid digestion prior to analysis by inductively coupled plasma-optical emission spectrometry. This includes several commodities like coffee, soy and cocoa beans.
3. Materials
Chemical/Solution | Description | Storage Location |
---|---|---|
Sample | Organic sample of commodity that will be analysed | / |
HNO3 (65%) | Highly concentrated Nitric acid (see Safety!) to digest sample | Room 227 ‘’Stock room’’, D2 |
Neutralizing agent (Chemizorb or NaHCO3) | Sodium bicarbonate (baking soda) or Chemizorb as neutralizing agent against accidental spills | Room 218b ‘’Main lab’’, A5 |
MilliQ water1 | Water of Millipore grade or equivalent used for general cleaning and more. Always see that several jerry-cans of Millipore water are available as a lot of it is used during the procedure. | Room 218b ‘’Main lab’’, A17 |
4. Equipment
Equipment | Description | Storage Location |
---|---|---|
PPE | Personal Protective Equipment, which includes safety glasses and a laboratory coat | Room 218b ‘’Main lab’’ |
Analytical Balance | Analytical Balance Sartorius CP 224 S that weighs samples accurately up to 4 decimal places | Room 218b ‘’Main lab’’, A7 |
Plastic spoon | Plastic spoon to transfer sample (if milled into powder sample) quantitatively into pressure vessels | Room 218b ‘’Main lab’’, A8 |
Pressure Vessels with vessel caps | PTFE Pressure Vessels and appropriate PTFE caps included with the Multiwave GO Plus microwave system | Room 218b ‘’Main lab’’, A8 |
Laminate gloves and acid-resistant sleeves | Safety gloves made of neoprene or better (e.g. butyl rubber). Standard nitrile gloves do NOT suffice to handle highly corrosive acids ( HNO3, >65%) | Room 218b ‘’Main lab’’, AV |
Protective apron | Apron to add an extra layer of protection against splashing and spilling of highly corrosive acid | Room 218b ‘’Main lab’’, AV |
Microwave system | A Multiwave GO Plus microwave system by Anton Paar (Figure 1) | Room 218b ‘’Main lab’’, A4 |
Dispenser | Acid resistant dispenser intended to efficiently dispense a correct volume of HNO3 | Room 218b ‘’Main lab’’, A5 |
Volumetric flask(s), 50 mL | Volumetric flask (50 mL) to dilute the digest after acid digestion.. This dilution is usually a prerequisite to analyze the digest with ICP-OES. | Room 218b ‘’Main lab’’, A15 |
Glass funnel | Glass funnel to catch rinse water | Room 218b ‘’Main lab’’, A15 |
5. Safety
5.1 H/P phrases HNO3 (65%)
5.2 Safety Summary: HNO3 (65%)
Inhalation
- Toxic if inhaled
- Corrosive to the respiratory tract
\=> work in fume hood
Skin
- Causes severe skin burns
\=> Wear laboratory coat, protective apron and safety gloves (laminate or neoprene)
Eyes
- Causes serious eye damage
\=> Wear safety glasses
Fire
-Oxidiser
\=> Keep away from heat sources
! Keep a neutralizing agent (Chemizorb or NaHCO3) nearby in case of accidental spills
5.3 Before starting digestion
5.3.1 Checklist
5.3.2 Place exhaust hose
The exhaust hose deducts the produced acid vapors safely from the instrument to your exhaust system. Multiwave GO Plus has to be connected to a fume hood with a minimum exhaust capacity of 100 m3/h. Make sure that the clip of the hose is tightened properly. To do so, fix the exhaust hose with the supplied hose clip to the flange of the exhaust of Multiwave GO Plus. Install the exhaust hose in a U shape (Figure 2, left) to avoid reflux of condensed acids. Carefully fix the open end of the exhaust hose with the delivered clip in the fume hood. Unscrew the metal rod and attach the exhaust hose (Figure 2, middle & right). While the instrument is working do not perform labor work in the same fume hood which is used for the exhaust hose of the instrument.
Figure 2: Correct attachment of exhaust hose to the fume hood in U shape
5.4 During digestion
5.4.1 General safety precautions
Do not lean against the instrument. Instead, it is recommended to keep a safe distance of approx. 1m from Multiwave GO Plus during operation.
During operation (microwave energy "ON") an electromagnetic field is generated by two transformers. People with pacemakers should keep a distance of at least 1m to the instrument during operation to ensure maximum safety. Keep the following items away from Multiwave GO Plus: credit cards or other cards with a magnetic read-write band. These might be damaged (e.g. if carried in the operator's pocket).
5.4.2 Power Failure
In case of power failure during the operation, please refer to the Instruction Manual and Safety Information or the Reference Guide of Multiwave GO Plus.
6. Sample Digestion
6.1 Preparing the pressure vessel HVT50
For coffee beans, soybeans and timber: weigh 0.5 g -1 g (Figure 3)
Figure 3: Weighing ball milled sample powder (right) directly into the pressure vessel
Table 1: volume of HNO3 to be added when using Organic A on the Multiwave GO Plus
Sample amount (g) | Acid volume (mL) |
---|---|
0-0.6 | 6 |
0.7 | 7 |
0.8 | 8 |
0.9 | 9 |
1 | 10 |
Figure 4: Correct way of closing the pressure vessel
6.2 Loading the rotor 12HVT50
Uni-vessel digestion
If only one pressure vessel, activate the single vessel mode in the run settings and load the vessel on position 1 (Figure 5). If a spark error (0x0024) occurs, check if acid venting took place, increase the amount of reagents, or increase the number of used vessels.
Figure 5: load patterns for the rotor
6.3 Running an experiment
The recommended temperature program for organic samples is ‘’Organic A’’. To create a new program, please refer to the Instruction Manual and Safety Information or the Reference Guide of Multiwave GO Plus
Figure 6: Start screen
Define the following settings:
1) Enter a "Run Name"
2) Select a "Method" e.g. preinstalled method ‘’Organic A’’. Move the desired list item into the red selection bar and tap \<Select>.
3) Optionally enter a "Run Note"
Tap \< > to confirm your settings.
Figure 7: Run screen
6.4 Viewing and exporting experiment results
Figure 8: How to open Data Memory and select your desired run
6.5 Unloading the rotor
Dry the lid and keep it separately from the rotor in the venting hood. The lid should not stay on the rotor, as condensed acid can corrode the top of the rotor.
Figure 9: Proper way of putting away the exhaust
7. Cleaning, maintenance and storage
Before you clean the instrument switch off the instrument and disconnect the mains plug.
(To remove oily residues, use alcohol/acetone but do NOT clean the screw caps in solvents like acetone! The O-rings will swell and can be destroyed)
Intense cleaning of pressure vessels after several runs:
1. Remove any visible impurities from the vessel body with water, a lab cleaning detergent and a soft brush.
2. Perform a cleaning run (see earlier).
3. Place the liner and vessels into a drying oven at 200 °C to 250 °C for at least 4 hours, or overnight. Any residues, which are not removed before heating the vessels in the drying oven will be burned-in and can be hardly removed afterwards.
4. Let the vessels cool down prior to first use.
Clean microwave chamber with damp cloth after each working day:
1. Remove the drive ring and the drip cup from the microwave cavity and clean with a damp cloth.
2. Clean the lid in the same way, also the contacting surfaces between lid and instrument.
3. Reassemble the parts and make sure they are completely dry before they are reused again.
Exchange IR sensor foil (if contaminated) + recalibrate IR sensor every month:
1. Remove the drive ring and drip cup from the cavity.
2. Remove the old IR sensor foil from the cavity (if contaminated). Remove the protective paper from the sticking area of the new IR sensor foil and align correctly (the ear of the foil should point inwards)
3. Stick the IR sensor foil directly onto the recess of the IR-sensor
4. Recalibrate:
1) Open the lid of Multiwave GO Plus
2) Remove the drive ring
3) Place the calibration unit into the two bores of the microwave cavity and connect the plug on the rear side of Multiwave GO Plus (Figure 10)
Figure 10: Correct placement of calibration unit
4) Tap ‘’Menu > Service > T-Calibration". The calibration is now performed.
5) Remove the calibration unit from the instrument.
Replace sealing foam of the rotor every 1-2 months
Cleaning after a vessel breakage \=> refer to Multiwave GO Plus Reference Guide
8. Waste disposal
9. Appendix
Acid | Volume Limitations | Advantages | Disadvantages |
---|---|---|---|
HNO3 | Between 3 mL and 15 mL | - Dissolves most metals | - Noble metals (Au, Pt, …) are not oxidized - Some metals (Al, Cr, …) are passivated |
HCl | Always add HNO3 to pure HCl. Aqua Regia is commonly used (HNO3 3:1 HCl). | -Complexation of several metals such as Fe(II), Fe (III), Sn (IV) and Ti (III) - Powerful in combination with HNO3 (Aqua Regia) - Adding HNO3 improves energy absorption - Stabilizes several elements | - Corrosive to vessels - AgCl, HgCl & TiCl are insoluble - PbCl2 is slightly soluble - absorbs microwave energy inefficiently |
H2O2 | Max. 6% of the reagent mixture + max. 2 mL per vessel | - Oxidizing power increases | - Can react explosively with organics |
HF | Max. 0.5 mL HF per 100 mg expected SiO2 | - Can dissolve Si - Complexation with H3BO3 (6 mL per 1 mL HF) resolubilizes insoluble elements + conserves ICP glassware | - Alkaline earth metals, lanthanides and actinides are insoluble - Toxic ! - Damages ICP glassware and Teflon |
H2SO4 | Max. 20% of the reagent mixture | - Increases reaction temperature at low pressure | - Stains + corrosion on vessels - BaSO4, SrSO4, PbSO4, CdSO4,... are insoluble |
H3PO4 | / | - Increases reaction temperature at low pressure | - Stains on vessels - MgPO4, CaPO4, … are insoluble |