Acid digestion of samples using a Multiwave GO Plus microwave system

1. Introduction
This standard operating procedure (SOP) describes a procedure for acid digestion of samples using the Multiwave GO Plus microwave system. A known weight of the sample is placed in the digestion vessel and digestion acid added. The digestion vessel is capped and subjected to microwave heating within a sealed pressurized chamber to achieve a high digestion temperature, then cooled down to room temperature. The digest liquor is quantitatively transferred into a plastic test tube and diluted with water to a known volume.

2. Scope
This procedure may be applied to any organic sample that is amenable to acid digestion prior to analysis by inductively coupled plasma-optical emission spectrometry. This includes several commodities like coffee, soy and cocoa beans.

3. Materials

Chemical/Solution Description Storage Location
Sample Organic sample of commodity that will be analysed /
HNO3 (65%) Highly concentrated Nitric acid (see Safety!) to digest sample Room 227 ‘’Stock room’’, D2
Neutralizing agent (Chemizorb or NaHCO3) Sodium bicarbonate (baking soda) or Chemizorb as neutralizing agent against accidental spills Room 218b ‘’Main lab’’, A5
MilliQ water1 Water of Millipore grade or equivalent used for general cleaning and more. Always see that several jerry-cans of Millipore water are available as a lot of it is used during the procedure. Room 218b ‘’Main lab’’, A17

4. Equipment

Equipment Description Storage Location
PPE Personal Protective Equipment, which includes safety glasses and a laboratory coat Room 218b ‘’Main lab’’
Analytical Balance Analytical Balance Sartorius CP 224 S that weighs samples accurately up to 4 decimal places Room 218b ‘’Main lab’’, A7
Plastic spoon Plastic spoon to transfer sample (if milled into powder sample) quantitatively into pressure vessels Room 218b ‘’Main lab’’, A8
Pressure Vessels with vessel caps PTFE Pressure Vessels and appropriate PTFE caps included with the Multiwave GO Plus microwave system Room 218b ‘’Main lab’’, A8
Laminate gloves and acid-resistant sleeves Safety gloves made of neoprene or better (e.g. butyl rubber). Standard nitrile gloves do NOT suffice to handle highly corrosive acids ( HNO3, >65%) Room 218b ‘’Main lab’’, AV
Protective apron Apron to add an extra layer of protection against splashing and spilling of highly corrosive acid Room 218b ‘’Main lab’’, AV
Microwave system A Multiwave GO Plus microwave system by Anton Paar (Figure 1) Room 218b ‘’Main lab’’, A4
Dispenser Acid resistant dispenser intended to efficiently dispense a correct volume of HNO3 Room 218b ‘’Main lab’’, A5
Volumetric flask(s), 50 mL Volumetric flask (50 mL) to dilute the digest after acid digestion.. This dilution is usually a prerequisite to analyze the digest with ICP-OES. Room 218b ‘’Main lab’’, A15
Glass funnel Glass funnel to catch rinse water Room 218b ‘’Main lab’’, A15

5. Safety

5.1 H/P phrases HNO3 (65%)

5.2 Safety Summary: HNO3 (65%)
Inhalation
- Toxic if inhaled
- Corrosive to the respiratory tract
\=> work in fume hood
Skin
- Causes severe skin burns
\=> Wear laboratory coat, protective apron and safety gloves (laminate or neoprene)
Eyes
- Causes serious eye damage
\=> Wear safety glasses
Fire
-Oxidiser
\=> Keep away from heat sources
! Keep a neutralizing agent (Chemizorb or NaHCO3) nearby in case of accidental spills

5.3 Before starting digestion

5.3.1 Checklist

5.3.2 Place exhaust hose
The exhaust hose deducts the produced acid vapors safely from the instrument to your exhaust system. Multiwave GO Plus has to be connected to a fume hood with a minimum exhaust capacity of 100 m3/h. Make sure that the clip of the hose is tightened properly. To do so, fix the exhaust hose with the supplied hose clip to the flange of the exhaust of Multiwave GO Plus. Install the exhaust hose in a U shape (Figure 2, left) to avoid reflux of condensed acids. Carefully fix the open end of the exhaust hose with the delivered clip in the fume hood. Unscrew the metal rod and attach the exhaust hose (Figure 2, middle & right). While the instrument is working do not perform labor work in the same fume hood which is used for the exhaust hose of the instrument.

Figure 2: Correct attachment of exhaust hose to the fume hood in U shape

5.4 During digestion

5.4.1 General safety precautions
Do not lean against the instrument. Instead, it is recommended to keep a safe distance of approx. 1m from Multiwave GO Plus during operation.

During operation (microwave energy "ON") an electromagnetic field is generated by two transformers. People with pacemakers should keep a distance of at least 1m to the instrument during operation to ensure maximum safety. Keep the following items away from Multiwave GO Plus: credit cards or other cards with a magnetic read-write band. These might be damaged (e.g. if carried in the operator's pocket).

5.4.2 Power Failure
In case of power failure during the operation, please refer to the Instruction Manual and Safety Information or the Reference Guide of Multiwave GO Plus.

6. Sample Digestion

6.1 Preparing the pressure vessel HVT50

  1. Make sure the pressure vessels are completely dry and clean
  2. Weigh the raw sample (or sample powder from ball mill2) into the vessel on an analytical balance (don’t forget to tare!):
  1. Repeat the weighing process for every vessel
  2. Add 2 mL MilliQ water to each vessel
  3. Place a warning sign at the door of the lab so other laborants know that pure acids will be handled
  4. Put on acid-resistant neoprene or laminate gloves AND sleeves.
  5. Add HNO3 to each vessel via a dispenser that is set to the correct volume. If using the generic method Organic A, add a volume of HNO3 according to Table 1. For a small number of vessels (1 or 3) use an acid amount of not less than 6 mL to provide enough mass for microwave absorption. For other acids, see Appendix I

Table 1: volume of HNO3 to be added when using Organic A on the Multiwave GO Plus

Sample amount (g) Acid volume (mL)
0-0.6 6
0.7 7
0.8 8
0.9 9
1 10
  1. Ensure that all vessel parts are completely dry: remove any acid or water droplets from the vessel surface. Ensure that no residues of the sample are sticking on the sealing and screwing area of the vessel body
  2. Place the seal onto the vessel and close the screw cap with your fingers until the spring holder and the spring holder insert are on the same level. The screw cap must be closed without overlap at the vent hole otherwise the rotor can be hot, even if cooling has supposedly finished and the lid lock has been released (Figure 4)

Figure 4: Correct way of closing the pressure vessel

  1. Place the close vessel into the rotor as described in 6.2
  2. Repeat for the other vessels

6.2 Loading the rotor 12HVT50

  1. Multi-vessel digestion
    Enter the vessels into the rotor. To ensure uniform temperature distribution during the run, the rotor has to be loaded symmetrically3. Follow the load patterns as described in Figure 5. If samples and blanks will be digested in the same run, mix them alternating in the rotor. If a spark error (0x0024) occurs, check if acid venting took place, increase the amount of reagents, or increase the number of used vessels.

Uni-vessel digestion
If only one pressure vessel, activate the single vessel mode in the run settings and load the vessel on position 1 (Figure 5). If a spark error (0x0024) occurs, check if acid venting took place, increase the amount of reagents, or increase the number of used vessels.

Figure 5: load patterns for the rotor

  1. Put the rotor lid onto the rotor
  2. Open the lid of Multiwave GO Plus and place the rotor onto the drive ring
  3. Close the lid of Multiwave GO Plus (you should hear a click sound)

6.3 Running an experiment
The recommended temperature program for organic samples is ‘’Organic A’’. To create a new program, please refer to the Instruction Manual and Safety Information or the Reference Guide of Multiwave GO Plus

  1. Make sure you remove your protective gloves so the machine will not be contaminated
  2. Switch on the instrument via the power button on the back of the machine. The Multiwave GO Plus automatically displays the start screen (Figure 6). Operate the touchscreen with gentle fingertip pressure only to ensure a long lifetime of the touchscreen.

Figure 6: Start screen

  1. Tap \< > "Create Run" on the start screen.
  2. Define the following settings:
    1) Enter a "Run Name"
    2) Select a "Method" e.g. preinstalled method ‘’Organic A’’. Move the desired list item into the red selection bar and tap \<Select>.
    3) Optionally enter a "Run Note"

  3. Tap \< > to confirm your settings.

  4. Check if Multi vessel (or Single vessel) mode is selected under \<>
  5. Tap \<Start> to start the run. Multiwave GO Plus starts processing the chosen method
    (Figure 7).

Figure 7: Run screen

  1. Wait until the run has finished. The results of the experiment are stored and can be viewed and/or exported onto a USB stick via the "Data Memory" dialog

6.4 Viewing and exporting experiment results

  1. Tap \<Menu> on the start screen and select "Menu > Data Memory". The data memory with all processed runs opens. Select a run by scrolling the red frame (Figure 8).

Figure 8: How to open Data Memory and select your desired run

  1. Tap \<Details> to open detailed information of your run
  2. Scroll the "Details" window to navigate through the run details.
  3. Tap \<Export> to export the run to a USB device

6.5 Unloading the rotor

  1. After the cooling has finished take the rotor from the oven (be careful: might still be hot!) and place it into a fume hood.
  2. Take off the lid and clean any traces of liquid on the vessels’ surface and the inside of the lid
  3. The vessels may still be under overpressure, thus, when opening the vessel in a fume hood point the vessel away from the body. Carefully and slowly open the vessel to release the residual pressure. Take care of possible foaming due to dissolved gases.
  4. Remove the seal from the pressure vessel. If the seal sticks in the pressure vessel, tilt the seal carefully into the cap to remove it easily.
  5. Transfer the content of the pressure vessel into a volumetric flask (50 mL). You can use a stand to hold the volumetric flask upright.
  6. Rinse the seals and vessels with MilliQ water and catch the rinse water using a glass funnel. Wash the funnel and dilute the digest to 50 mL with MilliQ water. Don’t forget to homogenise from time to time. You can use a piece of parafilm to close the flasks for homogenizing.
  7. Transfer the diluted digests into the labeled ICP-OES sample tubes and proceed with your task at hand.
  8. Turn off the microwave system and put the exhaust hose as shown in Figure 9.
  9. Clean and dry the rotor, vessels and instrument and inspect all parts for damage as described in 7. Cleaning, maintenance and storage.
  10. Dry the lid and keep it separately from the rotor in the venting hood. The lid should not stay on the rotor, as condensed acid can corrode the top of the rotor.

    Figure 9: Proper way of putting away the exhaust

7. Cleaning, maintenance and storage
Before you clean the instrument switch off the instrument and disconnect the mains plug.

(To remove oily residues, use alcohol/acetone but do NOT clean the screw caps in solvents like acetone! The O-rings will swell and can be destroyed)

  1. Rinse the vessels, screw caps and seals thoroughly with deionized water.
  2. If different types of samples are used, perform a cleaning run to avoid memory effects. A cleaning run goes as follows:
    1) Fill each vessel with at least 6 mL of a diluted digestion reagent mixture (e.g. 3 mL HNO3 and 3 mL H2O).
    2) Place the vessels into the rotor and place the rotor onto the drive ring
    3) Close the lid of Multiwave GO Plus and perform a cleaning run.
  3. All parts must be completely dry before they are reused again. Note that seals should always be stored with the same screw caps. Store pressure vessels on the drying rack and unscrewed, as the thread of the cap could be harmed.
    - Clean rotor with damp cloth after every run + check the sealing foam of the rotor for deformation or damage:
  4. Remove visible contaminations (e.g. reagent spills) with a damp cloth.
  5. Parts must be completely dry before they are reused again.
  6. Store the rotor outside of the microwave without the lid.
    - Clean the table top and water bottles as these might have been in contact with contaminated gloves. Don’t forget to clean your hands at the end too!

Figure 10: Correct placement of calibration unit

4) Tap ‘’Menu > Service > T-Calibration". The calibration is now performed.
5) Remove the calibration unit from the instrument.

8. Waste disposal

9. Appendix

  1. Different acids
    IMPORTANT: It is not recommended to digest samples with different acid mixtures in the same run. HNO3, HCl and H2O2 are suitable for digesting most environmental, agricultural and food samples. HF is required for total digestion with subsequent determination of the concentration of certain recalcitrant elements such as SI and Al, whereby the remaining HF can be complexed with H3BO3. H2SO4 and H2SO4 are only used for open-vessel digestions.
Acid Volume Limitations Advantages Disadvantages
HNO3 Between 3 mL and 15 mL - Dissolves most metals - Noble metals (Au, Pt, …) are not oxidized - Some metals (Al, Cr, …) are passivated
HCl Always add HNO3 to pure HCl. Aqua Regia is commonly used (HNO3 3:1 HCl). -Complexation of several metals such as Fe(II), Fe (III), Sn (IV) and Ti (III) - Powerful in combination with HNO3 (Aqua Regia) - Adding HNO3 improves energy absorption - Stabilizes several elements - Corrosive to vessels - AgCl, HgCl & TiCl are insoluble - PbCl2 is slightly soluble - absorbs microwave energy inefficiently
H2O2 Max. 6% of the reagent mixture + max. 2 mL per vessel - Oxidizing power increases - Can react explosively with organics
HF Max. 0.5 mL HF per 100 mg expected SiO2 - Can dissolve Si - Complexation with H3BO3 (6 mL per 1 mL HF) resolubilizes insoluble elements + conserves ICP glassware - Alkaline earth metals, lanthanides and actinides are insoluble - Toxic ! - Damages ICP glassware and Teflon
H2SO4 Max. 20% of the reagent mixture - Increases reaction temperature at low pressure - Stains + corrosion on vessels - BaSO4, SrSO4, PbSO4, CdSO4,... are insoluble
H3PO4 / - Increases reaction temperature at low pressure - Stains on vessels - MgPO4, CaPO4, … are insoluble

  1. Any time water is mentioned, we intend millipore grade or equivalent unless specified otherwise 

  2. If using milled sample i.e. sample powder, use a plastic spoon to transfer the sample powder into the pressure vessels quantitatively 

  3. This means the sample weights also have to be balanced! 

Book resources on time

Fume hood + Add event